Figure 1: AquaMax4000 microplate washer and LifeSep magnetic stand can be applied to MAGPIX multiple magnetic bead detection system
material:
- MAGPIX Calibration Kit ( Life Technology, Cat.#MPX-CAL-K25 )
- Phosphate buffer ( PBS )
- LifeSep 96F Magnetic Stand ( Dexter Magnetic Technology, Cat.#2501008-1 )
- 96- well fully transparent microplate ( Greiner )
- ® Velos TM  High Connotation System ( Molecular Devices )
method:
Step 1: Dilute the magnetic beads in PBS in a ratio of 1:50 , and add 100 ul of magnetic bead mixture to each well of a 96- well plate.
Step 2: Place the microplate on ImageXpress Velos before washing to obtain the number of beads.
Step 3: Set up the AquaMax microplate washer: create a new microplate parameter in the software, select Greiner 96 well plate, height input 24.2mm ( normal microplate height + 10mm LifeSep magnetic frame height ) , use the instrument Self-contained microplate depth detection function, which can accurately detect the depth of the microplate and save the detection data in the program.
Step 4: Use PBS to clean the microplate on the AquaMax4000 washer. After the panel is set up according to the following parameters, the best results will be obtained (effectively reducing the loss of magnetic beads during the cleaning process).
Aspiration
- Location: Edge
- Shampoo aspiration rate: medium speed
- Washing head falling speed: 3
- Height: 5mm
Packing
- Volume: 150ul
- Shampoo dispensing rate: 5
- Repeat 2 times from the first step
- Aspiration (set as above)
The green fluorescent signal was scanned using a 532 nm laser source with a ImageXpress Velos high- intensity system . After software analysis, the number of beads in each well before and after microplate cleaning was obtained.
Number of beads per hole in the 96- well plate before cleaning
The average number of beads in each hole = 615
Number of beads per hole in a 96- well plate after cleaning
The average number of beads in each hole = 585
Retention rate = 95.1%
Figure 2: Selecting a hole in the microplate to image the results before and after cleaning Â
Before cleaning  After cleaning
Select D5 hole  Select D5 hole
Number of beads: 633 Â Number of beads: 605
in conclusion:
Effectively reducing the loss rate of the magnetic beads during the cleaning process is a key factor in ensuring the success of the MAGPIX test. When the AquaMax microplate washer cleans the MAGPIX magnetic beads in the 96- well plate , use the LifeSep magnetic frame to perform the instrument as described above. After setting, more than 95% of the beads remain in the microplate. The AquaMax microplate washer has a high degree of auto-adjustment and shampoo interchangeability , ensuring flexibility in adapting to various test requirements.